TEK-AIR FLOWMETER - APEX-PREMIER SERIES - HIGH PERFORMANCE FUME HOOD MONITOR

 

The APEX Premier Monitor ensures a safer operating environment for lab personnel with the most versatile fume hood and room transfer velocity monitor available. This monitor is suited for use on all types of fume hoods including Constant Volume, Variable Volume and High Performance Low Flow Hoods. The APEX Premier Monitor can also be used to monitor the velocity of transfer air between critical pressure controlled environments.

Tek-Air Apex Premier

 

.The Apex Premier utilizes state of the art technology to provide the highest level of performance achievable in a velocity monitor. The result is unparalleled reliability and simplistic operation. The Heart of the Apex Premier is the Smart Sensor Module which is responsible for accurate, repeatable and instantaneous measurement of fume hood face or room transfer velocity. The velocity measurement is made by an ultra-sensitive micro-bridge mass airflow sensor with a response time of less than 3 milliseconds. The analog signal provided by the micro-bridge is digitized by a high-resolution A/D converter and then further enhanced and factory calibrated over the operating velocity range by an automated test & calibration system. The resulting velocity measurement of the calibrated Smart Sensor Module is an extremely accurate and repeatable signal suitable for use in all fume hood applications.

Applications
All types of fumehoods including high performance low flow hoods, variable volume and constant volume

Advantages
Microprocessor-based performance
Microbridge mass airflow sensor technology and advanced automated factory calibration
Real-time graphical display of face velocity
Internal menu driven setup and field calibration
Programmable mute button, output relays for Purge or Alarm, digital inputs for alarm or setback
Entire system settings viewable via one button
Password protected calibration configuration
Self-contained face velocity sensor

 

SPECIFICATION OF APEX-PREMIER :

DISPLAY
Type: Graphic LCD matrix touch screen 128 x 64 pixels
Viewing Area: 2.04” x 1.32”
Numeric Mode: Displays real time face velocity
Range: 40 to 500 FPM
Resolution: Selectable for 1 FPM or 5 FPM
Accuracy: +/- 10% or 10 FPM max.
Update Rate: Selectable for slow, medium or fast
Alpha Mode: Displays descriptive text indicating status
Graphical Mode: Displays face velocity trend line
Display Units: Imperial or Metric

AUDIBLE HORN
Sound Level: 84 dB at 1-foot, Configurable to be active or inactive
Operation: When active, horn sounds during alarm condition
Re-beep Mode: Provides reminder horn muted and alarm not cleared

ALARM CONTACTS
Type: Two independent SPST relays
Specifications: 1A @ 30 VDC, 0.5A @ 125 VAC (resistive load) each
Operation: Programmable to indicate purge or alarm condition
Note: analog output option, relay one is 5A @250VAC or 30VDC

ANALOG OUTPUT (OPTIONAL)
Type: Voltage output, proportional to face velocity measurements
Scaling: 0 to 10VDC adjustable in 0.1VDC increments, proportional
to 0 to 1000FPM adjustable in 1 fpm increments
Voltage Output: Source 20mA max. factory calibrated (300K ohm load)

DIGITAL INPUTS
Type: Two independent dry contact inputs
Operation: Each input is user programmable
Functions: Annunciate external alarm or setback mode,
Text Message: Provide selectable text messages to the display
Time Delay: Programmable to initiate after a user defined delay

PUSH BUTTONS
Mute: Silences horn, programmable active/inactive
Purge: Activates purge relay, programmable active/inactive
Scroll: Enables user to review entire monitor configuration

AGENCY APPROVALS
CE Approved, UL Listed

INPUT POWER
24 VAC +/- 20% 50/60 Hz @ 8VA or 15VDC +/-20% @ 500mA

LED INDICATORS
Type: 8mm diffused lens
Green: Indicates normal condition, flashes for warning condition
Red: Indicates alarm condition, flashes on mute
Yellow: Indicates purge button has been depressed

ENVIRONMENT
Temperature: Storage: 0°F to 150°F
Operating: 40°F to 120°F
Compensated: 60°F to 80°F
Humidity: Storage: 10% to 90% RH, non-condensing
Operating: 20% RH to 90% RH, non-condensing
Electromagnetic: Conforms to EMC Standards EN61326 Class A

VELOCITY SENSOR
Type: Micro-bridge mass flow sensor
Range: Bidirectional +/- 1000 fpm
Response Time: Typical 1 ms, maximum 3 ms
Accuracy: +/- 0.35% reading, includes repeatability & hysteresis
Overpressure: 25 PSI

REFERENCE PROBE
Internal: Integrated to display face, no external probe required
External: Optional, consult factory

HOOD PROBE
Type: 7/16” diameter, feed through bushing, press fit, 5’ tubing
Material: Polyethylene

TUBING
Type: .170” ID ¼” OD Clear
Material: Ester based polyurethane

PHYSICAL CHARACTERISTICS
Size: Front bezel; 3.35” W x 5.1”H x 0.5”D
Rear Enclosure; 2.5” W x 4.25”H x 3.5”D
Weight: 1 lb.
Materials: Front bezel; ABS
Rear enclosure; 20 ga. sheet metal

 